Difference between revisions of "Lifefix log"

From DIDEAS Wiki
Jump to: navigation, search
m (LF #1 LOG)
m
Line 3: Line 3:
  
 
=LF #1 LOG=
 
=LF #1 LOG=
 +
*10-02-09 02:10 - started [[LF_TEST_103]]
 +
*10-01-09 23:00 - estimated 600 passes of [[LF_TEST_103]] during code development
 +
*09-30-09 - LF rebuilt with 2 loadwashers
 
*09-21-09 16:44 - replace charge amp USB serial cable w/isolated USB serial cable.  GND isolation verified, however, otherwise requires system testing
 
*09-21-09 16:44 - replace charge amp USB serial cable w/isolated USB serial cable.  GND isolation verified, however, otherwise requires system testing
 
*09-09-19 22:02 - complete 9520 passes of [[LF_TEST102]], stopped due to loadwasher bolt break, nut outside
 
*09-09-19 22:02 - complete 9520 passes of [[LF_TEST102]], stopped due to loadwasher bolt break, nut outside

Revision as of 06:10, 2 October 2009

PF Users Navigation:

Edit

LF #1 LOG

  • 10-02-09 02:10 - started LF_TEST_103
  • 10-01-09 23:00 - estimated 600 passes of LF_TEST_103 during code development
  • 09-30-09 - LF rebuilt with 2 loadwashers
  • 09-21-09 16:44 - replace charge amp USB serial cable w/isolated USB serial cable. GND isolation verified, however, otherwise requires system testing
  • 09-09-19 22:02 - complete 9520 passes of LF_TEST102, stopped due to loadwasher bolt break, nut outside
  • 09-09-19 16:13 - start with LF_TEST102
  • 09-09-19 16:12 - complete 217 passes of LF_TEST102
  • 09-09-19 16:02 - complete 582 passes of LF_TEST102
  • 09-09-19 15:30 - complete 2175 passes of LF_TEST102, stopped for code development,
  • 09-09-19 14:10 - start with LF_TEST102
  • 09-09-19 09:55 - completed 13812 passes of LF_TEST102 , stopped due to false positive belt break
  • 09-09-19 02:00 - replace MPJA 20amp supply w/GW Instek. MPJA supply appears to be damaged from 9-17 test, which is the likely cause of the supply voltage drooping.
  • 09-09-19 01:37 - start with LF_TEST102
  • 09-09-17 10:00 - restarted but stopped due to broken bolt after 44 cycles
  • 09-09-17 09:00 - supply for this experiment had dropped to 15 volts.
  • 09-09-17 09:00 - stopped due to annoying neighbors.
  • 09-09-17 04:40 - started 10000 passes
  • 09-09-17 04:30 - started 237 passes
  • 09-09-17 03:30 - ran a few hundred cycles during experiments.
  • 09-09-17 03:00 - 3 incidences of max plantar, and re-phased the motor current
  • 09-09-11 23:53 - completed 4969 passes of LF_TEST_101 stops to belt break, ball nut on floor, load washer hanging
  • 09-09-11 21:00 - started LF_TEST_101
  • 09-09-11 14:00 - LW tightened
  • 09-09-11 03:00 - notice that LW has slipped - is no longer pre-loaded.
  • 09-09-11 00:10 - during belt stress test, LF runs away and max plantar
  • 09-09-10 23:30 - replaced broken belt; three ball screw balls fall out (from yesterday's over extension)
  • 09-09-10 02:30 - complete ~3600 passes of [[LF_TEST100], stopped due to broken belt
  • 09-09-10 00:30 - started LF_TEST100
  • 09-09-10 20:00 - resonance experiments of inside only - excessive dors
  • 09-09-10 16:00 - resonance experiments of full LF w/Kristin